This short and simple video shows you how to change settings in standby mode of Thopaz+
Changeable settings are:
- Pressure: Choose to max. -10 kPa (-100 cmH2O, -100 mbar, -80 mmHg).
- Pressure unit: Choose between kPa, cmH2O, mmHg, mbar.
- Measured pressure: Choose between hide and show (only relevant in the air leak history). A higher measured pressure than set pressure can be an indication that the patient is able to regulate the negative pressure in the inter-pleural space.
- Fluid time 1: Choose between 24 h / 12 h / 8 h / 4 h / OFF. Shows the total amount of collected fluid over the last X hours.
- Fluid time 2: Choose between 2 h / 1 h / 0.5 h (30 min) / 0.25 h (15 min)/OFF. Shows the total amount of collected fluid over the last X hours/minutes.
- Fluid alarm: Choose between 500 ml / 5min, 15min or 60min; 400 ml / 5min, 15min or 60min; 300 ml / 5min, 15min or 60min; 200 ml / 5min, 15min or 60min; 100 ml / 5min, 15min or 60min; 50 ml / 5min, 15min or 60min/OFF. The Fluid alarm 407 appears if Thopaz+ detected a higher fluid collection in the last 5 min, 15min or 60min than the desired parameter.
- Air leak history: Choose between hide and show. This history can be displayed on the main display.
- Language: Choose your language.
- Save as standard: Choose if you want to save the new parameter as a standard
This short and simple video shows you how to change settings in standby mode of Thopaz+
Changeable settings are:
- Pressure: Choose to max. -10 kPa (-100 cmH2O, -100 mbar, -80 mmHg).
- Pressure unit: Choose between kPa, cmH2O, mmHg, mbar.
- Measured pressure: Choose between hide and show (only relevant in the air leak history). A higher measured pressure than set pressure can be an indication that the patient is able to regulate the negative pressure in the inter-pleural space.
- Fluid time 1: Choose between 24 h / 12 h / 8 h / 4 h / OFF. Shows the total amount of collected fluid over the last X hours.
- Fluid time 2: Choose between 2 h / 1 h / 0.5 h (30 min) / 0.25 h (15 min)/OFF. Shows the total amount of collected fluid over the last X hours/minutes.
- Fluid alarm: Choose between 500 ml / 5min, 15min or 60min; 400 ml / 5min, 15min or 60min; 300 ml / 5min, 15min or 60min; 200 ml / 5min, 15min or 60min; 100 ml / 5min, 15min or 60min; 50 ml / 5min, 15min or 60min/OFF. The Fluid alarm 407 appears if Thopaz+ detected a higher fluid collection in the last 5 min, 15min or 60min than the desired parameter.
- Air leak history: Choose between hide and show. This history can be displayed on the main display.
- Language: Choose your language.
- Save as standard: Choose if you want to save the new parameter as a standard