How to Choose a Thermocouple (with Practical Examples)

preview_player
Показать описание

▶ You can read the full post here

⌚Timestamps:
00:00 - Intro
00:15 - I) Range and accuracy
02:28 - II) Styles of thermocouples
04:41 - A) Cold junction compensation
06:20 - B) Remote temperature monitoring
08:10 - Examples

=============================

There are many different types of thermocouples. Most manufacturers have selection guides to help you decide what to purchase.

In addition, most reputable manufacturers will also have technical support specialists available that will walk you through a series of questions to help select just the right type of thermocouple for your application.

You’ll need to consider things like the range of temperatures you are trying to measure, and the accuracy you would like to have. This will help narrow down the two different metal wires that you need to use. The type of thermocouple is based on the type of metals used in the sensing wires.

A Type-K thermocouple is made from Chromel and Alumel alloy wires. It can be used to measure temperature from around -330 °F to over +2460 °F.

You need to keep in mind that the accuracy of a Type-K thermocouple may only be about ±5 °F across the entire range. Sometimes that’s fine, but sometimes more accuracy is needed.

If you were looking at a cryogenic application that needed more accuracy, but you didn’t need to worry about high temperatures, you could consider a Type-T thermocouple.

A Type-T thermocouple is made from a copper wire and a Copper-Nickel wire. Type-T thermocouples are usually accurate to within a degree or two... so that makes them about twice as accurate as Type-K thermocouples.

Type-T thermocouples can typically measure even lower than -330 °F but the upper end of the range is usually just over 600 °F. They are more accurate but have a more limited measurement range.

Many applications for temperature measurement can get away with just using thermocouple wire with the ends twisted together.

In cases where a little more protection is desired, the wires are encased in a probe, which is just some kind of protective sheath and some insulating material to help protect the wires.

Thermocouple probes come in Ungrounded, Grounded, and Exposed Junction style. The style of thermocouple probe you select will also be based on your application.

The thermocouple signal is based on the difference in temperature between the hot and cold junctions. But we need to make a small signal correction to convert that temperature difference into an absolute temperature.

Fortunately, most measurement instruments can perform cold junction compensation... either with standard built-in equipment or with an optional cold junction compensation add-on.

Cold junction compensation components measure the temperature of the metal at the cold junction wiring connections and then make a signal calculation correction.

If you need to extend the wiring of the thermocouple, you must use something called a thermocouple extension wire in order to reduce the amount of error.

A thermocouple extension wire is constructed of the same types of metal as the thermocouple, so when you need to extend a circuit, a thermocouple extension wire of the same type as the thermocouple must be used.

Similarly, if you need to use terminal blocks to connect the wiring, the terminal blocks must be constructed of the same types of metal as the individual wires.

It is not recommended to run thermocouple extension wire over a long distance due to the sensitivity of the signal to electrical noise.

If you are trying to go more than 50 to 100 ft, you need to consider using a temperature transmitter that converts the millivolt signal over to another signal type such as 4-20 mA.

If you have multiple temperature measurements to make in a remote location, a remote I/O rack from your PLC might end up being a good option.

=============================

=============================

Missed our most recent videos? Watch them here:

=============================

=============================

To stay up to date with our last videos, make sure to subscribe to this YouTube channel:

=============================

=============================

#RealPars #Thermocouple #Industry
Рекомендации по теме
Комментарии
Автор

This is the first time in my career understanding what is transmitter, thank you realpurs

thinkgloballyactlocally
Автор

This is what i want, recently i joined a comapany dealing with these type of product. Really thakyou for sharing this video

Twocats_Coperation
Автор

I was your student. Thank you you I got many technical knowledge

abdulgaffarmohammedshafee
Автор

Can we get a video on remote i/o units please. Great video.

khanyithegreat
Автор

Suggestion: Please mention the temperature ranges in both F & C as surely you do now have a global audience & not just a North American centric audience.

tanmayduragkar
Автор

@9:44 I think the Profinet cable should go to the PLC controller, not the PLC AI module. What do you think Realpars?

bitebonumbere
Автор

This video explains more about temperature control measurement.

saviournartey
Автор

Now a days you get Thermocouple Temperature Amplifier Switch (Universal amplifier PR4116) which converts analog value to digital value.

shivaghuge
Автор

Nice Video, request for pH Sensor Please

chickenz
Автор

How long should the probe be? And does this affect the result

Shanesxperiments
Автор

Bro i need thermo couple for my oven accuracy i need +_2 i have j type with fuji controller but it gives 12 degree diffrence when it reaches 150 i want accuracy i have nee thermocouple wire and controller still this problem happened dont know what to do

funfire