Heathkit Oscilloscope O-12 Restoration & Repair 2 - Component testing

preview_player
Показать описание
Testing of the resistors and doing redlines of the schematics on the Heathkit Oscilloscope model O12 including a mistake found either in the unit or the schematic. Also the ordering of components needed to continue this operation. To speed things up I decided to just order all caps to be replaced as part of the repair and restoration of this unit.
Рекомендации по теме
Комментарии
Автор

If you look at that schematic again, the voltages after the 100 ohm resistor are in the 300's. The 450 written there is the capacitor voltages.

YouAllKnowBob
Автор

I was watching another video on this scope. The guy found the same missing resister. He found that the assembly instructions did not mention the 100 ohm resister . He concluded that the schematic was an error.

jggmobile
Автор

I use Mouser for all replacement parts.

tubeDude
Автор

Did you save the parts list? I’d love to have it!